| 10 | 1. Replace Single Board Computer. |
| 103 | 1. Replace Single Board Computer. |
| 210 | 1. Clean/Replace Syringe Sensor assembly. |
| 300 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean front end assembly. |
| 4. Reseat cable connections to Main Unit. |
| 5. Replace Single Board Computer. |
| 301 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
| 311 | 1. Replace head cable. |
| 2. Clean/Replace Flex Ring Assembly. |
| 3. Clean front end assembly. |
| 4. Replace Interface board in head. |
| 5. Replace Single Board Computer. |
| 6. Clean/Replace Syringe Sensor assembly. |
| 7. Calibrate/Replace potentiometer. |
| 8. Reseat Display board in head. |
| 9. Clean Interface card in head. |
| 314 | 1. Reseat display screen and connectors in monitor. |
| 2. Replace touch screen. |
| 3. Replace Single Board Computer in monitor. |
| 322 | 1. Replace Single Board Computer in monitor. |
| 411 | 1. Clean/Replace Syringe Sensor assembly. |
| 867 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate or replace 48 volt power supply in main unit. |
| 1004 | 1. Replace keypad in head. |
| 2. Clean/Replace Display board in head. |
| 3. Clean Display board in head. |
| 4. Clean docking sensor (piston tip). |
| 5. Clean/Replace Syringe Sensor assembly. |
| 6. Reseat cable connector on back of monitor. |
| 7. Verify brightness control button on back of monitor is not stuck. |
| 8. Verify Stop switch on injector head is not stuck. |
| 9. Replace Interface board in head. |
| 10. Replace Single Board Computer in head. |
| 1306 | 1. Reseat head cable to extension cable. |
| 2. Replace Single Board Computer in head. |
| 1309 | 1. Reseat cable connections. |
| 1311 | 1. Replace Single Board Computer in head. |
| 1319 | 1. Replace head cable. |
| 2. Reseat cable connections to Main Unit. |
| 3. Replace ethernet hub in Main Unit. |
| 4. Replace Main Unit. |
| 5. Replace Single Board Computer in monitor. |
| 1420 | 1. Replace Single Board Computer in head. |
| 1506 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate/Replace potentiometer. |
| 3. Replace Single Board Computer in head. |
| 1805 | 1. Calibrate/Replace potentiometer. |
| 2. Replace Single Board Computer in head. |
| 1806 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate/Replace potentiometer. |
| 3. Replace motor. |
| 4. Replace Single Board Computer in head. |
| 1811 | 1. Clean docking sensor (piston tip). |
| 2. Reseat head cable to extension cable. |
| 3. Clean interface card and pressure sensor connector in head. |
| 1815 | 1. Clean Interface card in head. |
| 2. Clean Power Drive card in head. |
| 1819 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in head. |
| 1906 | 1. Clean/Replace Flex Ring assembly. |
| 2104 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Check connections to Interface card in head. |
| 4. Check connections to Syringe Sensor assembly. |
| 5. Replace Interface board in head. |
| 2105 | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Check connections to Interface card in head. |
| 4. Check connections to Syringe Sensor assembly. |
| 2106 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Interface board in head. |
| 2108 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
| 2113 | 1. Clean/Replace Syringe Sensor assembly (most likely B side). |
| 2114 | 1. Clean/Replace Syringe Sensor assembly (most likely B side). |
| 2115 | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Calibrate/Replace potentiometer. |
| 4. Reseat head cable to extension cable. |
| 5. Replace Single Board Computer in head. |
| 2116 | 1. Clean docking sensor (piston tip). |
| 2120 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Interface board in head. |
| 3. Replace Single Board Computer in head. |
| 2121 | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 2201 | 1. Reseat head cable to extension cable. |
| 2209 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| 2210 | 1. Clean docking sensor (piston tip). |
| 2211 | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean/Replace Flex Ring assembly. |
| 2212 | 1. Clean/Replace Flex Ring assembly. |
| 2600 | 1. Clean front end assembly. |
| 2. Clean handswitch connector in the back of the head. |
| 3. Replace handswitch. |
| 4. Clean/Replace Flex Ring assembly. |
| 5. Replace Interface board in head. |
| 2601 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Single Board Computer in head. |
| 2800 | 1. Replace Main Unit. |
| 2. Replace Single Board Computer in head. |
| 3003 | 1. Replace Power Drive card in head. |
| 2. Check head cable coming from the main unit. |
| 3. Calibrate or replace 48 volt power supply in main unit. |
| 4. Replace Interface board in Main Unit. |
| 3009 | 1. Replace Single Board Computer in head. |
| 3013 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
| 3015 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
| 3016 | 1. Reseat connectors and Single Board Computer in head. |
| 2. Replace motor. |
| 3. Replace Single Board Computer in head. |
| 3017 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
| 3026 | 1. Calibrate or replace 48 volt power supply in main unit. |
| 2. Replace Power Drive card in head. |
| 3. Reseat extension cable connections. |
| 4. Replace head cable. |
| 5. Replace extension cable. |
| 6. Replace Interface board in Main Unit. |
| 7. Replace Main Unit. |
| 8. Replace Single Board Computer in head. |
| 3028 | 1. Calibrate or replace 48 volt power supply in main unit. |
| 2. Reseat extension cable connections. |
| 3. Replace extension cable. |
| 4. Replace head cable. |
| 3104 | 1. Clean front end assembly. |
| 3207 | 1. Replace Interface board in head. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
| 4. Clean Interface card in head. |
| 5. Clean Power Drive card in head. |
| 3401 | 1. Clean/Replace Syringe Sensor assembly. |
| 3407 | 1. Replace Interface board in head. |
| 2L3 on head display. | 1. Clean/Replace Syringe Sensor assembly. |
| 888 on head display. | 1. Clean/Replace Display Overlay board on head. |
| 2. Loose connection to Interface card in head. |
| 3. Replace Interface board in head. |
| Buttons on injector head not responding. | 1. Clean/Replace Display board on head. |
| Cannot arm injector from head. | 1. Clean/Replace Display board on head. |
| Communication error. | 1. Reseat cable connections. |
| 2. Replace head cable. |
| 3. Replace extension cable. |
| Dashes displayed on injector head. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| Detecting syringe when none is present. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| Display on head is locked up. | 1. Clean/Replace Display board on head. |
| Docking error. | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Flex Ring assembly. |
| Docking pin on the end of the piston tip is stuck. | 1. Clean docking sensor (piston tip). |
| Encoder Out of Tolerance. | 1. Replace motor. |
| Foreign Syringe detected. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| FRB | 1. Clean/Replace Flex Ring assembly. |
| 2. Reseat extension cable to head cable. |
| Green arm light will not come on. | 1. Replace Interface board in head. |
| Handswitch shorted. | 1. Inspect/Replace handswitch. |
| Head display blank. | 1. Clean/Replace Display board on head. |
| Heat maintainer not working. | 1. Check heat maintainers connection to Interface card inside head. |
| Heater blinking. | 1. Replace syringe heater. |
| HOVER | 1. Clean/Replace Display board on head. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Interface board in head. |
| 5. Replace Single Board Computer in head. |
| HSW | 1. Inspect/Replace handswitch. |
| Injector keeps rebooting itself. | 1. Replace Single Board Computer. |
| Injector powering down. | 1. Inspect cable connection to monitor. |
| Invalid syringe error. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| ISI module not working. | 1. Replace ISI module. |
| 2. Reboot ISI module. |
| 3. Reset power to the ISI module. |
| 4. Remove "Hold" phase in the protocol. |
| 5. Turn on ISI option in menu. |
| 6. Replace ISI interface cable. |
| 7. Replace Interface board on CT scanner. |
| LL3 on injector head display. | 1. Clean/Replace Syringe Sensor assembly. |
| Monitor display black. | 1. Replace Interface board in head. |
| 2. Replace bad ribbon cable going from Single Board Computer to display panel in monitor. |
| 3. Replace Invertor board in monitor. |
| 4. Replace Single Board Computer in monitor. |
| Monitor display blank. | 1. Reseat cable connections in monitor. |
| 2. Reseat cable connections to Main Unit. |
| 3. Replace Invertor board in monitor. |
| 4. Replace Single Board Computer in monitor. |
| Monitor display blinking. | 1. Reseat cable connections in monitor. |
| Monitor display fading. | 1. Replace Single Board Computer. |
| Monitor display has lines across it. | 1. Reseat cable connections in monitor. |
| 2. Replace Single Board Computer in monitor. |
| Monitor display image creeping up from bottom. | 1. Replace Single Board Computer in monitor. |
| Monitor display white. | 1. Reseat display ribbon cable in monitor. |
| 2. Check/Calibrate 48 volt power supply in main unit. |
| 3. Replace ethernet hub in Main Unit. |
| 4. Replace Interface board in Main Unit. |
| 5. Replace Single Board Computer ribbon cable in monitor. |
| 6. Replace Single Board Computer in monitor. |
| Monitor keeps turning off. | 1. Reseat cable connections in monitor. |
| 2. Replace Single Board Computer in monitor. |
| Monitor locking up. | 1. Reseat cable connections to Main Unit. |
| 2. Replace Single Board Computer in head. |
| No pressure displayed on pesssure graph. | 1. Replace Interface board in head. |
| Noisy piston movement. | 1. Clean piston and internal O-ring. |
| 2. Remove manual knobs and lens cap and tighten nut on piston gear. |
| Piston does not stop during auto-docking. | 1. Clean piston tip. |
| Piston making a clicking noise when moving. | 1. Remove manual knobs and lens cap and tighten nut on piston gear. |
| Piston not auto-docking to syringe when installed. | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Calibrate/Replace potentiometer. |
| 5. Replace Interface board in head. |
| Piston not auto-retracting when syringe is removed. | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
| Piston not grabbing syringe plunger during reverse. | 1. Clean docking sensor (piston tip). |
| 2. Adjust allen screws on piston tip. |
| 3. Piston rubber boot rotated out of position. |
| Piston not releasing syringe. | 1. Piston rubber boot out of position. |
| Piston stops in the middle of the injection. | 1. Clean docking sensor (piston tip). |
| 2. Replace motor. |
| POTA | 1. Calibrate/Replace potentiometer. |
| POTB | 1. Calibrate/Replace potentiometer. |
| POWR | 1. Replace Power Drive card in head (see also error 3026/3028). |
| Pressure limiting. | 1. A disposable adaptor at the injection site is causing the pressure limit. |
| Protocol Lock light on head is always on. | 1. Clean/Replace Display board on head. |
| Purge button on injector head not working. | 1. Clean/Replace Display board on head. |
| Reset button on monitor is not functioning. | 1. Replace Button interface and cable in monitor. |
| Some of the buttons on the injector head are hard to press. | 1. Clean/Replace Display board on head. |
| Start button on injector head not working. | 1. Clean/Replace Display board on head. |
| Stop button on injector head not working. | 1. Clean/Replace Display board on head. |
| SW | 1. inspect/Replace Head Cable. |
| 2. Replace Single Board Computer. |
| Syringe difficult to remove. | 1. Clean piston tip and slightly loosen the set screw at the bottom of the piston tip. |
| 1. Clean/Replace Flex Ring assembly. |
| Syringe interface error. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in injector head. |
| Syringe not detected. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in injector head. |