10 | 1. Replace Single Board Computer. |
103 | 1. Replace Single Board Computer. |
210 | 1. Clean/Replace Syringe Sensor assembly. |
300 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean front end assembly. |
| 4. Reseat cable connections to Main Unit. |
| 5. Replace Single Board Computer. |
301 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
311 | 1. Replace head cable. |
| 2. Clean/Replace Flex Ring Assembly. |
| 3. Clean front end assembly. |
| 4. Replace Interface board in head. |
| 5. Replace Single Board Computer. |
| 6. Clean/Replace Syringe Sensor assembly. |
| 7. Calibrate/Replace potentiometer. |
| 8. Reseat Display board in head. |
| 9. Clean Interface card in head. |
314 | 1. Reseat display screen and connectors in monitor. |
| 2. Replace touch screen. |
| 3. Replace Single Board Computer in monitor. |
322 | 1. Replace Single Board Computer in monitor. |
411 | 1. Clean/Replace Syringe Sensor assembly. |
867 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate or replace 48 volt power supply in main unit. |
1004 | 1. Replace keypad in head. |
| 2. Clean/Replace Display board in head. |
| 3. Clean Display board in head. |
| 4. Clean docking sensor (piston tip). |
| 5. Clean/Replace Syringe Sensor assembly. |
| 6. Reseat cable connector on back of monitor. |
| 7. Verify brightness control button on back of monitor is not stuck. |
| 8. Verify Stop switch on injector head is not stuck. |
| 9. Replace Interface board in head. |
| 10. Replace Single Board Computer in head. |
1306 | 1. Reseat head cable to extension cable. |
| 2. Replace Single Board Computer in head. |
1309 | 1. Reseat cable connections. |
1311 | 1. Replace Single Board Computer in head. |
1319 | 1. Replace head cable. |
| 2. Reseat cable connections to Main Unit. |
| 3. Replace ethernet hub in Main Unit. |
| 4. Replace Main Unit. |
| 5. Replace Single Board Computer in monitor. |
1420 | 1. Replace Single Board Computer in head. |
1506 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate/Replace potentiometer. |
| 3. Replace Single Board Computer in head. |
1805 | 1. Calibrate/Replace potentiometer. |
| 2. Replace Single Board Computer in head. |
1806 | 1. Clean/Replace Flex Ring assembly. |
| 2. Calibrate/Replace potentiometer. |
| 3. Replace motor. |
| 4. Replace Single Board Computer in head. |
1811 | 1. Clean docking sensor (piston tip). |
| 2. Reseat head cable to extension cable. |
| 3. Clean interface card and pressure sensor connector in head. |
1815 | 1. Clean Interface card in head. |
| 2. Clean Power Drive card in head. |
1819 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in head. |
1906 | 1. Clean/Replace Flex Ring assembly. |
2104 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Check connections to Interface card in head. |
| 4. Check connections to Syringe Sensor assembly. |
| 5. Replace Interface board in head. |
2105 | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Check connections to Interface card in head. |
| 4. Check connections to Syringe Sensor assembly. |
2106 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Interface board in head. |
2108 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
2113 | 1. Clean/Replace Syringe Sensor assembly (most likely B side). |
2114 | 1. Clean/Replace Syringe Sensor assembly (most likely B side). |
2115 | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Calibrate/Replace potentiometer. |
| 4. Reseat head cable to extension cable. |
| 5. Replace Single Board Computer in head. |
2116 | 1. Clean docking sensor (piston tip). |
2120 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Interface board in head. |
| 3. Replace Single Board Computer in head. |
2121 | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
2201 | 1. Reseat head cable to extension cable. |
2209 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
2210 | 1. Clean docking sensor (piston tip). |
2211 | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean/Replace Flex Ring assembly. |
2212 | 1. Clean/Replace Flex Ring assembly. |
2600 | 1. Clean front end assembly. |
| 2. Clean handswitch connector in the back of the head. |
| 3. Replace handswitch. |
| 4. Clean/Replace Flex Ring assembly. |
| 5. Replace Interface board in head. |
2601 | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Single Board Computer in head. |
2800 | 1. Replace Main Unit. |
| 2. Replace Single Board Computer in head. |
3003 | 1. Replace Power Drive card in head. |
| 2. Check head cable coming from the main unit. |
| 3. Calibrate or replace 48 volt power supply in main unit. |
| 4. Replace Interface board in Main Unit. |
3009 | 1. Replace Single Board Computer in head. |
3013 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
3015 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
3016 | 1. Reseat connectors and Single Board Computer in head. |
| 2. Replace motor. |
| 3. Replace Single Board Computer in head. |
3017 | 1. Clean/Replace Flex Ring assembly. |
| 2. Replace Single Board Computer in head. |
3026 | 1. Calibrate or replace 48 volt power supply in main unit. |
| 2. Replace Power Drive card in head. |
| 3. Reseat extension cable connections. |
| 4. Replace head cable. |
| 5. Replace extension cable. |
| 6. Replace Interface board in Main Unit. |
| 7. Replace Main Unit. |
| 8. Replace Single Board Computer in head. |
3028 | 1. Calibrate or replace 48 volt power supply in main unit. |
| 2. Reseat extension cable connections. |
| 3. Replace extension cable. |
| 4. Replace head cable. |
3104 | 1. Clean front end assembly. |
3207 | 1. Replace Interface board in head. |
| 2. Clean docking sensor (piston tip). |
| 3. Clean/Replace Flex Ring assembly. |
| 4. Clean Interface card in head. |
| 5. Clean Power Drive card in head. |
3401 | 1. Clean/Replace Syringe Sensor assembly. |
3407 | 1. Replace Interface board in head. |
2L3 on head display. | 1. Clean/Replace Syringe Sensor assembly. |
888 on head display. | 1. Clean/Replace Display Overlay board on head. |
| 2. Loose connection to Interface card in head. |
| 3. Replace Interface board in head. |
Buttons on injector head not responding. | 1. Clean/Replace Display board on head. |
Cannot arm injector from head. | 1. Clean/Replace Display board on head. |
Communication error. | 1. Reseat cable connections. |
| 2. Replace head cable. |
| 3. Replace extension cable. |
Dashes displayed on injector head. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Clean docking sensor (piston tip). |
Detecting syringe when none is present. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
Display on head is locked up. | 1. Clean/Replace Display board on head. |
Docking error. | 1. Clean docking sensor (piston tip). |
| 2. Clean/Replace Flex Ring assembly. |
Docking pin on the end of the piston tip is stuck. | 1. Clean docking sensor (piston tip). |
Encoder Out of Tolerance. | 1. Replace motor. |
Foreign Syringe detected. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
FRB | 1. Clean/Replace Flex Ring assembly. |
| 2. Reseat extension cable to head cable. |
Green arm light will not come on. | 1. Replace Interface board in head. |
Handswitch shorted. | 1. Inspect/Replace handswitch. |
Head display blank. | 1. Clean/Replace Display board on head. |
Heat maintainer not working. | 1. Check heat maintainers connection to Interface card inside head. |
Heater blinking. | 1. Replace syringe heater. |
HOVER | 1. Clean/Replace Display board on head. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Replace Interface board in head. |
| 5. Replace Single Board Computer in head. |
HSW | 1. Inspect/Replace handswitch. |
Injector keeps rebooting itself. | 1. Replace Single Board Computer. |
Injector powering down. | 1. Inspect cable connection to monitor. |
Invalid syringe error. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
ISI module not working. | 1. Replace ISI module. |
| 2. Reboot ISI module. |
| 3. Reset power to the ISI module. |
| 4. Remove "Hold" phase in the protocol. |
| 5. Turn on ISI option in menu. |
| 6. Replace ISI interface cable. |
| 7. Replace Interface board on CT scanner. |
LL3 on injector head display. | 1. Clean/Replace Syringe Sensor assembly. |
Monitor display black. | 1. Replace Interface board in head. |
| 2. Replace bad ribbon cable going from Single Board Computer to display panel in monitor. |
| 3. Replace Invertor board in monitor. |
| 4. Replace Single Board Computer in monitor. |
Monitor display blank. | 1. Reseat cable connections in monitor. |
| 2. Reseat cable connections to Main Unit. |
| 3. Replace Invertor board in monitor. |
| 4. Replace Single Board Computer in monitor. |
Monitor display blinking. | 1. Reseat cable connections in monitor. |
Monitor display fading. | 1. Replace Single Board Computer. |
Monitor display has lines across it. | 1. Reseat cable connections in monitor. |
| 2. Replace Single Board Computer in monitor. |
Monitor display image creeping up from bottom. | 1. Replace Single Board Computer in monitor. |
Monitor display white. | 1. Reseat display ribbon cable in monitor. |
| 2. Check/Calibrate 48 volt power supply in main unit. |
| 3. Replace ethernet hub in Main Unit. |
| 4. Replace Interface board in Main Unit. |
| 5. Replace Single Board Computer ribbon cable in monitor. |
| 6. Replace Single Board Computer in monitor. |
Monitor keeps turning off. | 1. Reseat cable connections in monitor. |
| 2. Replace Single Board Computer in monitor. |
Monitor locking up. | 1. Reseat cable connections to Main Unit. |
| 2. Replace Single Board Computer in head. |
No pressure displayed on pesssure graph. | 1. Replace Interface board in head. |
Noisy piston movement. | 1. Clean piston and internal O-ring. |
| 2. Remove manual knobs and lens cap and tighten nut on piston gear. |
Piston does not stop during auto-docking. | 1. Clean piston tip. |
Piston making a clicking noise when moving. | 1. Remove manual knobs and lens cap and tighten nut on piston gear. |
Piston not auto-docking to syringe when installed. | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
| 4. Calibrate/Replace potentiometer. |
| 5. Replace Interface board in head. |
Piston not auto-retracting when syringe is removed. | 1. Clean/Replace Flex Ring assembly. |
| 2. Clean/Replace Syringe Sensor assembly. |
| 3. Clean docking sensor (piston tip). |
Piston not grabbing syringe plunger during reverse. | 1. Clean docking sensor (piston tip). |
| 2. Adjust allen screws on piston tip. |
| 3. Piston rubber boot rotated out of position. |
Piston not releasing syringe. | 1. Piston rubber boot out of position. |
Piston stops in the middle of the injection. | 1. Clean docking sensor (piston tip). |
| 2. Replace motor. |
POTA | 1. Calibrate/Replace potentiometer. |
POTB | 1. Calibrate/Replace potentiometer. |
POWR | 1. Replace Power Drive card in head (see also error 3026/3028). |
Pressure limiting. | 1. A disposable adaptor at the injection site is causing the pressure limit. |
Protocol Lock light on head is always on. | 1. Clean/Replace Display board on head. |
Purge button on injector head not working. | 1. Clean/Replace Display board on head. |
Reset button on monitor is not functioning. | 1. Replace Button interface and cable in monitor. |
Some of the buttons on the injector head are hard to press. | 1. Clean/Replace Display board on head. |
Start button on injector head not working. | 1. Clean/Replace Display board on head. |
Stop button on injector head not working. | 1. Clean/Replace Display board on head. |
SW | 1. inspect/Replace Head Cable. |
| 2. Replace Single Board Computer. |
Syringe difficult to remove. | 1. Clean piston tip and slightly loosen the set screw at the bottom of the piston tip. |
| 1. Clean/Replace Flex Ring assembly. |
Syringe interface error. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in injector head. |
Syringe not detected. | 1. Clean/Replace Syringe Sensor assembly. |
| 2. Clean/Replace Flex Ring assembly. |
| 3. Replace Interface board in injector head. |