Medrad Stellant Errors

Error CodeSolution
101. Replace Single Board Computer.
1031. Replace Single Board Computer.
2101. Clean/Replace Syringe Sensor assembly.
3001. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Clean front end assembly.
4. Reseat cable connections to Main Unit.
5. Replace Single Board Computer.
3011. Clean/Replace Syringe Sensor assembly.
2. Clean docking sensor (piston tip).
3. Clean/Replace Flex Ring assembly.
3111. Replace head cable.
2. Clean/Replace Flex Ring Assembly.
3. Clean front end assembly.
4. Replace Interface board in head.
5. Replace Single Board Computer.
6. Clean/Replace Syringe Sensor assembly.
7. Calibrate/Replace potentiometer.
8. Reseat Display board in head.
9. Clean Interface card in head.
3141. Reseat display screen and connectors in monitor.
2. Replace touch screen.
3. Replace Single Board Computer in monitor.
3221. Replace Single Board Computer in monitor.
4111. Clean/Replace Syringe Sensor assembly.
8671. Clean/Replace Flex Ring assembly.
2. Calibrate or replace 48 volt power supply in main unit.
10041. Replace keypad in head.
2. Clean/Replace Display board in head.
3. Clean Display board in head.
4. Clean docking sensor (piston tip).
5. Clean/Replace Syringe Sensor assembly.
6. Reseat cable connector on back of monitor.
7. Verify brightness control button on back of monitor is not stuck.
8. Verify Stop switch on injector head is not stuck.
9. Replace Interface board in head.
10. Replace Single Board Computer in head.
13061. Reseat head cable to extension cable.
2. Replace Single Board Computer in head.
13091. Reseat cable connections.
13111. Replace Single Board Computer in head.
13191. Replace head cable.
2. Reseat cable connections to Main Unit.
3. Replace ethernet hub in Main Unit.
4. Replace Main Unit.
5. Replace Single Board Computer in monitor.
14201. Replace Single Board Computer in head.
15061. Clean/Replace Flex Ring assembly.
2. Calibrate/Replace potentiometer.
3. Replace Single Board Computer in head.
18051. Calibrate/Replace potentiometer.
2. Replace Single Board Computer in head.
18061. Clean/Replace Flex Ring assembly.
2. Calibrate/Replace potentiometer.
3. Replace motor.
4. Replace Single Board Computer in head.
18111. Clean docking sensor (piston tip).
2. Reseat head cable to extension cable.
3. Clean interface card and pressure sensor connector in head.
18151. Clean Interface card in head.
2. Clean Power Drive card in head.
18191. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Replace Interface board in head.
19061. Clean/Replace Flex Ring assembly.
21041. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Check connections to Interface card in head.
4. Check connections to Syringe Sensor assembly.
5. Replace Interface board in head.
21051. Clean docking sensor (piston tip).
2. Clean/Replace Flex Ring assembly.
3. Check connections to Interface card in head.
4. Check connections to Syringe Sensor assembly.
21061. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Clean docking sensor (piston tip).
4. Replace Interface board in head.
21081. Clean/Replace Syringe Sensor assembly.
2. Clean docking sensor (piston tip).
3. Clean/Replace Flex Ring assembly.
21131. Clean/Replace Syringe Sensor assembly (most likely B side).
21141. Clean/Replace Syringe Sensor assembly (most likely B side).
21151. Clean/Replace Flex Ring assembly.
2. Clean/Replace Syringe Sensor assembly.
3. Calibrate/Replace potentiometer.
4. Reseat head cable to extension cable.
5. Replace Single Board Computer in head.
21161. Clean docking sensor (piston tip).
21201. Clean/Replace Flex Ring assembly.
2. Replace Interface board in head.
3. Replace Single Board Computer in head.
21211. Clean/Replace Flex Ring assembly.
2. Clean/Replace Syringe Sensor assembly.
22011. Reseat head cable to extension cable.
22091. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Clean docking sensor (piston tip).
22101. Clean docking sensor (piston tip).
22111. Clean docking sensor (piston tip).
2. Clean/Replace Syringe Sensor assembly.
3. Clean/Replace Flex Ring assembly.
22121. Clean/Replace Flex Ring assembly.
26001. Clean front end assembly.
2. Clean handswitch connector in the back of the head.
3. Replace handswitch.
4. Clean/Replace Flex Ring assembly.
5. Replace Interface board in head.
26011. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Clean docking sensor (piston tip).
4. Replace Single Board Computer in head.
28001. Replace Main Unit.
2. Replace Single Board Computer in head.
30031. Replace Power Drive card in head.
2. Check head cable coming from the main unit.
3. Calibrate or replace 48 volt power supply in main unit.
4. Replace Interface board in Main Unit.
30091. Replace Single Board Computer in head.
30131. Clean/Replace Flex Ring assembly.
2. Replace Single Board Computer in head.
30151. Clean/Replace Flex Ring assembly.
2. Replace Single Board Computer in head.
30161. Reseat connectors and Single Board Computer in head.
2. Replace motor.
3. Replace Single Board Computer in head.
30171. Clean/Replace Flex Ring assembly.
2. Replace Single Board Computer in head.
30261. Calibrate or replace 48 volt power supply in main unit.
2. Replace Power Drive card in head.
3. Reseat extension cable connections.
4. Replace head cable.
5. Replace extension cable.
6. Replace Interface board in Main Unit.
7. Replace Main Unit.
8. Replace Single Board Computer in head.
30281. Calibrate or replace 48 volt power supply in main unit.
2. Reseat extension cable connections.
3. Replace extension cable.
4. Replace head cable.
31041. Clean front end assembly.
32071. Replace Interface board in head.
2. Clean docking sensor (piston tip).
3. Clean/Replace Flex Ring assembly.
4. Clean Interface card in head.
5. Clean Power Drive card in head.
34011. Clean/Replace Syringe Sensor assembly.
34071. Replace Interface board in head.
2L3 on head display.1. Clean/Replace Syringe Sensor assembly.
888 on head display.1. Clean/Replace Display Overlay board on head.
2. Loose connection to Interface card in head.
3. Replace Interface board in head.
Buttons on injector head not responding.1. Clean/Replace Display board on head.
Cannot arm injector from head.1. Clean/Replace Display board on head.
Communication error.1. Reseat cable connections.
2. Replace head cable.
3. Replace extension cable.
Dashes displayed on injector head.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Clean docking sensor (piston tip).
Detecting syringe when none is present.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
Display on head is locked up.1. Clean/Replace Display board on head.
Docking error.1. Clean docking sensor (piston tip).
2. Clean/Replace Flex Ring assembly.
Docking pin on the end of the piston tip is stuck.1. Clean docking sensor (piston tip).
Encoder Out of Tolerance.1. Replace motor.
Foreign Syringe detected.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
FRB1. Clean/Replace Flex Ring assembly.
2. Reseat extension cable to head cable.
Green arm light will not come on.1. Replace Interface board in head.
Handswitch shorted.1. Inspect/Replace handswitch.
Head display blank.1. Clean/Replace Display board on head.
Heat maintainer not working.1. Check heat maintainers connection to Interface card inside head.
Heater blinking.1. Replace syringe heater.
HOVER1. Clean/Replace Display board on head.
2. Clean/Replace Syringe Sensor assembly.
3. Clean docking sensor (piston tip).
4. Replace Interface board in head.
5. Replace Single Board Computer in head.
HSW1. Inspect/Replace handswitch.
Injector keeps rebooting itself.1. Replace Single Board Computer.
Injector powering down.1. Inspect cable connection to monitor.
Invalid syringe error.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
ISI module not working.1. Replace ISI module.
2. Reboot ISI module.
3. Reset power to the ISI module.
4. Remove "Hold" phase in the protocol.
5. Turn on ISI option in menu.
6. Replace ISI interface cable.
7. Replace Interface board on CT scanner.
LL3 on injector head display.1. Clean/Replace Syringe Sensor assembly.
Monitor display black.1. Replace Interface board in head.
2. Replace bad ribbon cable going from Single Board Computer to display panel in monitor.
3. Replace Invertor board in monitor.
4. Replace Single Board Computer in monitor.
Monitor display blank.1. Reseat cable connections in monitor.
2. Reseat cable connections to Main Unit.
3. Replace Invertor board in monitor.
4. Replace Single Board Computer in monitor.
Monitor display blinking.1. Reseat cable connections in monitor.
Monitor display fading.1. Replace Single Board Computer.
Monitor display has lines across it.1. Reseat cable connections in monitor.
2. Replace Single Board Computer in monitor.
Monitor display image creeping up from bottom.1. Replace Single Board Computer in monitor.
Monitor display white.1. Reseat display ribbon cable in monitor.
2. Check/Calibrate 48 volt power supply in main unit.
3. Replace ethernet hub in Main Unit.
4. Replace Interface board in Main Unit.
5. Replace Single Board Computer ribbon cable in monitor.
6. Replace Single Board Computer in monitor.
Monitor keeps turning off.1. Reseat cable connections in monitor.
2. Replace Single Board Computer in monitor.
Monitor locking up.1. Reseat cable connections to Main Unit.
2. Replace Single Board Computer in head.
No pressure displayed on pesssure graph.1. Replace Interface board in head.
Noisy piston movement.1. Clean piston and internal O-ring.
2. Remove manual knobs and lens cap and tighten nut on piston gear.
Piston does not stop during auto-docking.1. Clean piston tip.
Piston making a clicking noise when moving.1. Remove manual knobs and lens cap and tighten nut on piston gear.
Piston not auto-docking to syringe when installed.1. Clean/Replace Flex Ring assembly.
2. Clean/Replace Syringe Sensor assembly.
3. Clean docking sensor (piston tip).
4. Calibrate/Replace potentiometer.
5. Replace Interface board in head.
Piston not auto-retracting when syringe is removed.1. Clean/Replace Flex Ring assembly.
2. Clean/Replace Syringe Sensor assembly.
3. Clean docking sensor (piston tip).
Piston not grabbing syringe plunger during reverse.1. Clean docking sensor (piston tip).
2. Adjust allen screws on piston tip.
3. Piston rubber boot rotated out of position.
Piston not releasing syringe.1. Piston rubber boot out of position.
Piston stops in the middle of the injection.1. Clean docking sensor (piston tip).
2. Replace motor.
POTA1. Calibrate/Replace potentiometer.
POTB1. Calibrate/Replace potentiometer.
POWR1. Replace Power Drive card in head (see also error 3026/3028).
Pressure limiting.1. A disposable adaptor at the injection site is causing the pressure limit.
Protocol Lock light on head is always on.1. Clean/Replace Display board on head.
Purge button on injector head not working.1. Clean/Replace Display board on head.
Reset button on monitor is not functioning.1. Replace Button interface and cable in monitor.
Some of the buttons on the injector head are hard to press.1. Clean/Replace Display board on head.
Start button on injector head not working.1. Clean/Replace Display board on head.
Stop button on injector head not working.1. Clean/Replace Display board on head.
SW1. inspect/Replace Head Cable.
2. Replace Single Board Computer.
Syringe difficult to remove.1. Clean piston tip and slightly loosen the set screw at the bottom of the piston tip.
1. Clean/Replace Flex Ring assembly.
Syringe interface error.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Replace Interface board in injector head.
Syringe not detected.1. Clean/Replace Syringe Sensor assembly.
2. Clean/Replace Flex Ring assembly.
3. Replace Interface board in injector head.